Automatic Isothermal Adsorption Analyzer

Automatic Isothermal Adsorption Analyzer
Product Detailed

Isothermal Adsorption Process

1. Lifting Samples

Pretreatment of the coal sample is accurately weighed and quickly into the model. Sample weight: about 2kg. Industrial analysis implements are in accordance with GB/T212991 of coal sample industrial analysis method.

2. Air Tightness Inspection

2.1 Air Inflation

Filled with Helium to system, the pressure is higher than highest pressure 1MPa of isothermal adsorption experiment. System temperature is adjusted to reservoir temperature. Continuously observe for the system seal condition if good, go to next experiment.

3. Free Space Volume Measurement

3.1 Temperature

System temperature is adjusted to reservior temperature.

3.2 Inflation

Open the cylinder, inflate Helium gas to system, adjust standard pressure valve, and then close the standard room valve.

3.3 Data Acquisition

Open the valve of standard room and sample cylinder, collect the data after pressure balance, do the volume experiment twice. The allowable error is ±0.1cm3.

3.4 Calculate the coal volume and get the blank volume of sample room

4. Isothermal Adsorption Experiment

4.1 Ensure Experiment Pressure

The highest experiment pressure is set up to 1.2 times of reservoir . The minimum pressure is 8MPa.

4.2 Experiment Pressure Spot Distribution

When highest pressure is lower than 8MPa, choose 6 experiment spots; Highest pressure between 8MPa to 12MPa, choose 7 experiment spots; Highest pressure >12MPa, choose 8 experiment spots.

4.3 Inflation

Open regulated valve and standard room Valve, inflate methane to system and adjust the standard room pressure to ideal pressure.

4.4 Data Collection

After get the ideal pressure and stable temperature, start the data collection of the isothermal adsorption experiment program.

4.5 Ensure Time of Adsorption Balance

More than 12 hours at least

4.6 Repeat

Step 3-5 until the end of experiment.

Main Performance Index and Parameter

Parameter

Temperature

Room Tem -100℃

Operation Pressure

40MPa

Maximum Pressure

70MPa

Sample Cylinder, Reference Cylinder

 

Sample Particle Size

≤60um

Sample Weight

200g

Reference Cylinder Volume

150ml

Reference Cylinder Quantities

6pcs OD: 70mm Length: 330mm

Sampler Cylinder Volume

150ml

Material

Stainless Steel

Pressure Seal

Specified Metal Ring Seal

Stressor

Gas Booster Pump

Pressure Device

Air Compressor

Thermostat

 

Oil Bathpan

60L*3pcs

Oil Bathpan Size

500mm(Length)*480mm(width)*240mm(depth)

Temperature Setting

Independent Zones

Oil Bathpan Circulation

Internal Installation

Constant Temperature Oil

with it

Pressure Acquisition

 

Pressure Sensor

Three Wire Pressure Sensor

Pressure Range

0-40MPa

Related Pressure

40MPa

One-piece Brometer

Precision±0.1%

Quantity

12pcs

Output

3mv/v

Temperature Acquisition

 

Temperature Detection System

Temperature Sensor±0.1℃

Multi-channel Digital Converter Precision

16

Gassource

 

Aggregate Supply

3

Gas Input

Methane Helium Carbon Dioxide

Pressure

20MPa

Stability

3%

Electric Gas Regulator

 

Aggregate Supply

3

Pressure

40MPa

Stability

1%

Input

0-10v

Booster Pressure

0.6MPa

Computer Control

 

Computer

Basic Frequency0.3GHz

Internal Storage 1GB, Hard Disk: 160GB

Auto Control

Computer

System Software

FINESORB-ISO Data Acquisition Software

FINESORB-ISO Adsorb/Desorb Data Processing Software

Multicomponent Gas Adsorb/Desorb Data Processing

Software