Precious Metal Dispersion Analyzer
Main Property
1. A variety of workflows- TPD, TPR, TPO, TPSR, Pulse Titration, Pretreatment and BET Specific Surface Area
2. Multiple Independent Gassource- In addition to carrier gas and reaction gas, each for independent control and mixes in scale.
3. Multiple Gas Joints - 3 for carrier gas and 3 for reaction gas
4. Steam Generator – It is equipped with a saturated steam generator and convenient to use reaction gas like liquid NH3.
5. Can be connected to GC or MS – It can be connected to the gas chromatography (GC) or mass spectroscopy (MS).
6. Auxiliary Detector – The instrument can output a voltage signal which is convenient to connect gas auxiliary detector, such as FID.
7. Material Structure - Stainless steel piping, seal for fluorine rubber, ask company for special custom if applied in special environment.
Precise Control
1. Flow Control – Using mass flow controller (MFC) software control, the flow is adjustable between 10 to 100sccm and software controls gas coefficient.
2. High Temperature Furnace – The highest temperature is 1000℃, and the stove heating rate can be adjustable from 1℃/min to 30℃/min.
3. Air Refrigeration – Open (shell type) furnace can be quickly cooling and optional refrigerator achieving rapid refrigeration.
4. Optional Cryotrap – The temperature can be rising from -100℃.
5. Temperature Control – Piping, steam generator, stove and TCD temperature can all be controlled by digit.
6. Thermal Conductivity Cell Detector – 4-arm rhenium tungsten TCD has extremely high linear, accuracy, sensitivity and stability.
7. Very small dead volume - Small volume valve and 3 mm pipe can reduce dead volume, thus reduce the deviation of the peak.
Easy Operation
1. Sample Support – U type pipe of large size can be convenient for sample loading and cleaning, and can add powder, rod or shot sample.
2. Sample Loading – Sample tube free of taking down is convenient to load samples.
Safety
1. Heating Pipeline – All the pipeline, valve and liquid carburetor are heated to prevent condensation.
2. Safety – Independent furnace temperature thermal protection, sealing pressure reducing valve, current circuit breaker, fuse, the system design has timely security protection.
Parameter
Temperature System |
|
||||||||
Range: |
70 to 1100 °C with CoolTec Cooler™ |
|
|||||||
option, Ambient to 1100 °C (without CoolTec Cooler ) |
|
||||||||
Selection: |
Digitally set, 1 °C increments |
|
|||||||
Ramp Rates: |
Up to 60 °C per minute within Ambient to 600 °C range |
|
|||||||
Up to 30 °C per minute from 600 to 800 °C range |
|
||||||||
Up to 10 °C per minute from 800 to 1100 °C range |
|
||||||||
|
|
||||||||
Gases |
|
||||||||
Loop (Analysis): |
H2, CO, O2, N2O, NH3, |
|
|||||||
vapors such as pyridine, water, etc. |
|
||||||||
Carrier: |
He, Ar, and other gases |
|
|||||||
Preparation: |
H2, O2, He, Ar, and others |
||||||||
|
|
||||||||
Gas Flow Rate |
|
||||||||
All Mass Flow Controllers (MFCs) |
|
||||||||
Manual Control: |
0 to 100 sccm |
|
|||||||
Automatic Analysis: |
5 to 85 sccm Analysis: |
|
|||||||
|
|
||||||||
Gas Delivery |
|
||||||||
Inlet Ports: |
3 each for carrier gas, |
|
|||||||
treatment (analysis) gas |
|
||||||||
Temperature Control: |
Internal gas lines and valves heated up to 150 °C |
|
|||||||
|
|
||||||||
Sample Tube |
|
||||||||
Type: |
Fused quartz FlowThru sample tubes, |
|
|||||||
for use up to 1100 °C, accepts powders and |
|
||||||||
pellets up to 8 mm in diameter |
|
||||||||
|
|
||||||||
Electrical |
|
||||||||
Voltage: |
85 to 265 VAC |
|
|||||||
Frequency: |
50/60 Hz |
|
|||||||
Power: |
900 VA, operating, max. |
|
|||||||
|
|
||||||||