Zhejiang Finetec Instruments Co., Ltd

Precious Metal Dispersion Analyzer

Precious Metal Dispersion Analyzer
Product Detailed

Main Property

1. A variety of workflows- TPD, TPR, TPO, TPSR, Pulse Titration, Pretreatment and BET Specific Surface Area

2. Multiple Independent Gassource- In addition to carrier gas and reaction gas, each for independent control and mixes in scale.

3. Multiple Gas Joints - 3 for carrier gas and 3 for reaction gas

4. Steam Generator – It is equipped with a saturated steam generator and convenient to use reaction gas like liquid NH3.

5. Can be connected to GC or MS – It can be connected to the gas chromatography (GC) or mass spectroscopy (MS).

6. Auxiliary Detector – The instrument can output a voltage signal which is convenient to connect gas auxiliary detector, such as FID.

7. Material Structure - Stainless steel piping, seal for fluorine rubber, ask company for special custom if applied in special environment.

 

Precise Control

1. Flow Control – Using mass flow controller (MFC) software control, the flow is adjustable between 10 to 100sccm and software controls gas coefficient.

2. High Temperature Furnace – The highest temperature is 1000℃, and the stove heating rate can be adjustable from 1℃/min to 30℃/min.

3. Air Refrigeration – Open (shell type) furnace can be quickly cooling and optional refrigerator achieving rapid refrigeration.

4. Optional Cryotrap – The temperature can be rising from -100℃.

5. Temperature Control – Piping, steam generator, stove and TCD temperature can all be controlled by digit.

6. Thermal Conductivity Cell Detector – 4-arm rhenium tungsten TCD has extremely high linear, accuracy, sensitivity and stability.

7. Very small dead volume - Small volume valve and 3 mm pipe can reduce dead volume, thus reduce the deviation of the peak.

 

Easy Operation

1. Sample Support – U type pipe of large size can be convenient for sample loading and cleaning, and can add powder, rod or shot sample.

2. Sample Loading – Sample tube free of taking down is convenient to load samples. 

 

 

Safety

1. Heating Pipeline – All the pipeline, valve and liquid carburetor are heated to prevent condensation.

2. Safety – Independent furnace temperature thermal protection, sealing pressure reducing valve, current circuit breaker, fuse, the system design has timely security protection.

Parameter
  



Copyright Notice @ 2008-2022 B2BAGE Limited and/or its subsidiaries and licensors. All rights reserved.

Temperature System

 

Range:

70 to 1100 °C with CoolTec Cooler™

 

option, Ambient to 1100 °C (without CoolTec Cooler )

 

Selection:

Digitally set, 1 °C increments

 

Ramp Rates:

Up to 60 °C per minute within Ambient to 600 °C range

 

Up to 30 °C per minute from 600 to 800 °C range

 

Up to 10 °C per minute from 800 to 1100 °C range

 

 

 

Gases

 

Loop (Analysis):

H2, CO, O2, N2O, NH3,

 

vapors such as pyridine, water, etc.

 

Carrier:

He, Ar, and other gases

 

Preparation:

H2, O2, He, Ar, and others

 
 

 

 

Gas Flow Rate

 

All Mass Flow Controllers (MFCs)

 

Manual Control:

0 to 100 sccm

 

Automatic Analysis:

5 to 85 sccm Analysis:

 

 

 

Gas Delivery

 

Inlet Ports:

3 each for carrier gas,

 

treatment (analysis) gas

 

Temperature Control:

Internal gas lines and valves heated up to 150 °C

 

 

 

Sample Tube

 

Type:

Fused quartz FlowThru sample tubes,

 

for use up to 1100 °C, accepts powders and

 

pellets up to 8 mm in diameter

 

 

 

Electrical

 

Voltage:

85 to 265 VAC

 

Frequency:

50/60 Hz

 

Power:

900 VA, operating, max.