Zhejiang Finetec Instruments Co., Ltd

Autosampler 2200

Autosampler 2200
Product Detailed

Main Property

1. Concise appearance, small size, simple and convenient placement.

2. Compatible with many types of gas chromatography both at home and abroad.

3. 15 sample disks

4. High precision imported stepping motor, which improves the precision and accuracy of analysis to a large extent.

5. Automatic analysis for sample injection, competent for long time continuous operation.

6. Strong anti-interference ability, meeting various industrial environment demand and suitable to various laboratory environment.

7. 5 kinds of predefined methods: standard shunt/no shunt, sample injection mode in the standard column, high purity sample, vicidity sample (such as cyclohexanol) and easy volatile sample (pentane), meet the needs of analysis of most samples.

8. 5 kinds of custom methods. You can set methods of your own needs combining sample characteristics with specific method of sample injection, optimizing the results of your analysis.

9. Prior injection mode. In the automatic operation process of sample sequence, important sample which needs temporary analysis can be inserted, and at the same time ensure that the sequence of sample injection does not interrupt, greatly improving the flexibility of analysis.

10. Manual injection mode. The sampler can be directly used after setting a few simple parameters for analysis, no need to set multifarious analysis sequence for sample.

11. Selection for a variety of automatic injection mode: conventional sample injection, continuous sample injection and PTV sample injection.

12. Support double ports for sample injection. Each port can choose their own analysis method, sample, injection quantity and injection delay.

13.  Graphical dynamic operation interface, intuitive conveniently observe the current running status of FAS2200.

14. Sample injection process intervened artificially. The current sample running sequence can be forced to the end.

15. Intelligent error checking. The system automatically inspects the parameter users set, effectively avoiding misoperation for sampler.

16. Support switching in both Chinese and English.


Parameter

 

Injection Port

2 Ports(realize simultaneous injection)

Injection Mode

3 kinds(Routine Injection, Continuous Injection、PTV Injection)

PTV Max Time Interval

PTV Max time interval0~65536 S

Delay between two injector

0.5 - 10 Min

Max Time Interval of Injection

65536S

Injection Quantity

0.1 - 500 uL 

Internal Quantity

0 - 490 uL 

Solvent Quantity Columns

0 - 500 uL

Solvent Cleaning

solvent I,II,III

Solvent Cleaning before Injection

0 – 99times

Solvent Cleaning

after Injection

0 – 99times

Sample Cleaning before Injection

0 – 99 times

Exhaust Bubble

0 – 99times

Injection Depth

0 - 100%(Relative to the length of the needle)

Speed of Sampling

0.1 - 500 uL/sec

Speed of Injection

0.1 - 500 uL/sec

Viscosity Delay

0 - 99sec

Delay before Injection

0 - 99sec

Delay after Injection

0 - 99sec

Repeatability

0.4% RSD on-cold column injection, 1uL, FID detector

Power

24 Vac

Communication

RS232 and TTL

Volume(W x H x D)

130 mm x 250 mm x 380 mm

Weight

5.6 kg

Temperature

5℃-40℃

Relative Humidity

0% - 80%

 



Copyright Notice @ 2008-2022 B2BAGE Limited and/or its subsidiaries and licensors. All rights reserved.